Tuesday, February 7, 2012

X-Ray Diffractometry XRD

 X-Ray Diffractometry (XRD):
Principle:  In this technique the primary X-rays are made to fall on the sample substance
under study. Because of its wave nature, like light waves, it gets diffracted to a certain angle.
This angle of diffraction, which differs from that of the incident beam, will give the
information regarding the crystal nature of the substance. The wavelength of the X-rays can
be varied for the application by using a grating plate.

Instrumentation :  It consists of X-ray tube for the  source, monochromator and a rotating
detector.



Applications : The diffraction of X-rays is a good tool to study the nature of the crystalline
substances. In crystals the ions or molecules are arranged in well-defined positions in planes
in three dimensions. The impinging X-rays are reflected by each crystal plane. Since the
spacing between the atoms and hence the planes can’t be same or identical for any two
chemical substances, this technique provides vital information regarding the arrangement of
atoms and the spacing in between them and also to find out the chemical compositions of
crystalline substances. The sample under study can be of either a thin layer of crystal or in a
powder form. Since, the power of a diffracted beam is dependent on the quantity of the
corresponding crystalline substance, it is  also possible to carry out quantitative
determinations.

X-Ray Photo-emission Spectrometry XPS

 X-Ray Photo-emission Spectrometry (XPS): 
Principle: When a primary X-ray beam of precisely known energy impinges on sample 
atoms, inner shell electrons are ejected and the energy of the ejected electrons  is measured. 
The difference in the energy of the impinging  X-ray and the ejected  electrons gives the 
binding energy (Eb) of the electron to the atom. Since,  this binding energy of the emitted 
electron depends on the energy of the electronic orbit and the element it can be used to 
identify the element involved. Further, the chemical form or environment of the atom affects 
the binding energy to a considerable extent to give rise to some chemical shift, which can be 
used to identify the valence state of the atom and its exact chemical form. This technique is 
mostly referred as Electron Spectroscopy for Chemical Analysis (ESCA).  
 
An associated process with this method is that when the electron is ejected from the inner 
orbital a vacancy is left with. Hence, another electron from the outer orbits may fall to fill the 
vacancy and by doing so emits X-ray fluorescence. The energy of this X-ray fluorescence is 
sometimes transferred to a second electron to make it to be ejected. This second electron thus 
emitted is termed as  Auger electron  and the method  Auger Spectroscopy  (after French 
Physicist  Pierre Auger).  Its applications are more or less similar to ESCA and both the 
methods are used in conjunction since in both cases the energies involved are similar.   
 
Instrumentation:  It consists of a radiation source for primary X-rays, monochromator, the 
energy analyzer (to resolve the electrons generated from the samples by energy) and detector 
to measure the intensity of the resolved electrons. The analysis is done in high vacuum. 



Applications : It is mainly used for surface analysis, especially in the qualitative identification 
of the elements in a sample. Based on the chemical shifts, the chemical environment around 
the atoms can also be estimated. This measurement is useful in determining the valence states 
of the atoms present in various moieties in a sample. Quantitative measurements can be made 
by determining the intensity of the ESCA lines of each element. 
 
Disadvantages : High vacuum is necessary for the system to avoid the low energy electrons to 
be collided with other impurities, which may result in low sensitivity. It is not possible to 
detect the impurities at the ppm or ppb levels. The whole instrumentation is highly 
complicated.

X-Ray Fluorescence Spectrometry XFS

X-Ray Fluorescence Spectrometry (XFS) : 
Principle: When a sample is placed in a beam of primary X-rays, part of it will be absorbed 
and the atoms get excited, by the ejection of electrons present in K and L shells. While 
relaxing they re-emit X-rays of characteristic wavelength. This re-emitted X-rays are called 
secondary or fluorescent X-rays  and hence the name for this technique. Since, the 
wavelength of the fluorescence is characteristic of the element being excited, measurement of 
the wavelength and intensity enables to carry out the qualitative and quantitative analyses. 

Instrumentation:  It comprises of a source for primary X-rays, collimators, analyzing crystal 
and detector.  



Applications:    It is one of the non-destructive methods in the elemental analysis of solid or 
liquid samples for major and minor constituents. Most of the elements in the periodic table, 
both metals and nonmetals, respond to this technique. Detection limit is between 10 to 100 
ppm. One of the significant uses of this method is in the medical field in identifying and 
determining the sulfur in protein.  


Disadvantages:  The sensitivity gets affected for elements with lower atomic numbers, 
particularly elements with atomic number  lower than 15 are difficult to analyze. The 
sensitivity is also limited by matrix absorption, secondary fluorescence and scattering of the 
particles. Instruments are often large, complicated and costly. 

Generation of X-Rays

Generation of X-Rays :  When metals, like copper, molybdenum, tungsten, etc., are 
bombarded directly with a stream of high energy electrons or radioactive particles, X-rays  
(wavelengths of order 0.1-100Å) are emitted because of the transitions involving K-Shell and 
L-Shell electrons. This can be simply expressed as follows: 

A cathode in the form of a metal wire when  electrically heated gives off electrons. If a 
positive voltage, in the form of an anode (target comprised of the metals mentioned above), is 
placed near these electrons, the electrons are accelerated toward the anode. Upon striking the 
anode, the electrons transfer their energy to the metallic surface, which then gives off X-ray 
radiation. This is referred as primary X-rays.  
The following is the schematic diagram for the process:



Note:  The wavelength of the emitted X-ray is characteristic of the element being 
bombarded. Hence with some modifications this process can be used as a tool for 
qualitative and quantitative elemental analysis by measuring the wavelength and emission 
intensity of the X-rays respectively. This forms the basis for Electron Probe Microanalysis!