Showing posts with label spectroscopy. Show all posts
Showing posts with label spectroscopy. Show all posts

Monday, January 23, 2012

Electron Spin Resonance Spectroscopy ESR

Electron Spin Resonance Spectroscopy (ESR):
The basic principle of electron spin resonance spectroscopy is  that, electrons always have a
spin and thus have a magnetic moment. Thus, the magnetic resonance theory applies to
electrons too like that of nuclei, as in NMR. Especially this technique is of high value when it
comes to the compounds which contain odd electrons, i.e. those substances which have
paramagnetic behaviour (if electrons are paired as in bonded  orbital then their mutual
spinning will cancel each other and there will be no response for the applied magnetic field,




whereas, if it is unpaired then it can align with the applied magnetic field and the feasibility 
of getting ESR spectra is higher). Thus, the  principle and the instrumentation are much 
similar to that of NMR technique. It is also referred as,  Electron Magnetic Resonance 
(EMR) or Electron Paramagnetic Resonance (EPR) spectroscopy. 

It is mostly used as a potential technique to study the formation and lifetime of free radicals, 
which are the major intermediates in most of the organic reactions. Another important 
application is in the estimation of trace amounts of paramagnetic ions, particularly in 
biological works like, Mn2+, Mg2+  etc. 

Nuclear Magnetic Resonance Spectroscopy NMR

Nuclear Magnetic Resonance Spectroscopy (NMR) : 
Principle :  In NMR substances absorb energy in the radio frequency region of the electro-
magnetic spectrum under influence of a strong magnetic field. It is a well known fact that the 
nuclei of the atoms bonded to each other in molecules spin on an axis like a top. Since nuclei 
are positively charged, this spin will create a  small magnetic field. If an external magnetic 
field is applied to these nuclei this magnetic  field will split into two energy levels. The 
energy difference is very small and corresponds to radiofrequency energy which is unique for 
every molecule and will give the information regarding the nature of the compounds and the 
presence of various functional groups and their environment.  

Since this technique is mostly measures the spinning of the hydrogen nuclei (almost all the 
organic compounds contain hydrogen atoms!), it is sometimes referred as Proton Magnetic 
Resonance (PMR) spectroscopy. 

Instrumentation :  The instrumentation for this technique includes powerful magnet, radio-
frequency signal generator, amplifier, detector, etc. The following is the outline of the 
instrument: 



Applications :  The application lies mostly in the identification and structural analysis of 
organic compounds and thus, it is mostly a tool for qualitative  analysis.  It gives valuable 
information regarding the position of the functional groups in a molecule and provides 
distinguished spectra for the isomer. Much  precise information on the structure of the 
compounds can be obtained using the same technique with other magnetic nuclei like C 13
O17
, the instrumentation being the same except that the sweep of the magnetic field is varied. 
  
Disadvantages: Very expensive and the instrumentation is complex and needs exceptional 
skills to operate. Its sensitivity ranges from moderate to poor, however, can get clear 
information using C13
 or O17 
NMR. The usage of the solvents  is limited and in most of the 
situations deuterated solvents are required. 

Microwave Spectroscopy

Microwave Spectroscopy
This technique is actually an extension to IR spectroscopy. Microwave region lies at the far 
infra-red region of the electromagnetic spectrum and its absorption by molecules give rise to  
change in the  rotational energies  of the molecules. 

In IR spectroscopy, the molecules are subjected to changes in vibrational energies; the energy required for making changes at rotational levels is lesser than that for vibrational levels. Though the principles are same to 
that of IR, the instrumentation is slightly different and it requires samples in gaseous state for 
the analysis. 



Its applications are limited to smaller and simpler molecules since larger 
molecules will have interactions between the rotational energy levels within the molecule 
through various bonds they have. Besides qualitative analysis, this technique can be applied for conformational analysis  of simpler compounds (study of stereo chemistry of the compounds). 

Fourier Transform Raman Spectroscopy FT-Raman

Fourier Transform Raman Spectroscopy (FT-Raman): 
Principle : This technique is complementary to FT-IR and is a scattering technique, whereby a 
laser beam (near-IR region) is directed to the sample and the scattered radiation is collected. 
Most of the scattered radiation has the same wavenumber as that of the incident laser beam, 
however a fraction will be having a different wavenumber. This is the Raman signal and 
characteristic of particular functional group. Raman spectroscopy finds applications in 
identifying organic compounds containing non-polar bonds such as carbon - carbon double 
bonds or aromatic rings (weak dipoles).  

Instrumentation : The instrumentation comprises of exciting laser normally in near-IR region,  
Rayleigh filter, beam splitter, detector, etc. Data collection and processing are akin to IR  
including the Fourier transformations.  


Applications : The applications are similar to FT-IR and gives useful information on the non-
polar bonds, i.e. bonds with null or reduced dipole moment. Water is a good solvent for FT-
Raman. 

Disadvantages : Signal strength is normally weak, and liquid samples give poor signals. Heat 
sensitive samples can’t be analyzed, since local heating will damage the samples. Dark 
colored samples can’t be analyzed. 

Fourier Transform Infrared Spectroscopy FT-IR

Fourier Transform Infrared Spectroscopy (FT-IR)
Principles :   It involves the absorption of electromagnetic radiation in the infrared region of 
the spectrum which results in changes in the vibrational energy of molecule. Since, usually 
all molecules will be having vibrations in the form of stretching, bending, etc., the absorbed 
energy will be utilised in changing the energy levels associated with them. It is a valuable 
and formidable tool in identifying organic compounds which have polar chemical bonds 
(such as OH, NH, CH, etc.) with good charge separation (strong dipoles).  

Instrumentation :  It was originally designed as a double beam spectrophotometer comprising 
IR source (red hot ceramic material), grating monochromator, thermocouple detector, cells 
made of either sodium chloride or potassium bromide materials, etc. In this process the light 
is dispersed by the monochromator. But, this type of basic design for IR measurements has 
been outdated. Instead a newer technique termed Fourier Transform-Infrared (FT-IR) has 
been in practice. This technique utilises a single beam of un-dispersed light and has the 
instrument components similar to the previous one. 

In FT-IR, the un-dispersed light beam is passed through the sample and the absorbances at all 
wavelengths are received at the detector simultaneously. A computerized mathematical 
manipulation (known as “Fourier Transform”) is performed on this data, to obtain absorption 
data for each and every wavelength. To perform this type of calculations interference of light 
pattern is required for which the FT-IR instrumentation contains two mirrors, one fixed and 
one moveable with a beam splitter in between them. Before scanning the sample a reference 
or a blank scanning is required. The following is the simplified design of the instrument: 




Applications :  It finds extensive use in the identification and structural analysis of organic 
compounds, natural products, polymers, etc. The presence of particular functional group in a 
given organic compound can be identified.  Since every functional group has unique 
vibrational energy, the IR spectra can be seen as their fingerprints. 

Disadvantages :  Samples containing mixture of substances can not be analysed. Since the 
sample holders and beam splitter, are made  of moisture sensitive materials like sodium 
chloride or potassium bromide (KBr), special cells are required for aqueous samples (e.g. 
KRS-5, ZnSe, etc.). Water is a bad solvent for IR spectral works.

Fluorometry Molecular Fluorescence


Fluorometry:  Molecular Fluorescence 
Principle :  This technique utilises the phenomenon  of molecular fluorescence, the theory 
behind this is exactly the same that has been discussed under atomic fluorescence but through 
the excitation of bonded electrons. Here, most often the irradiating light is in the range of 
ultraviolet and visible.  

Instrumentation :  The instrumental set-up comprises of a UV/Visible source, two 
monochromators, detector and recorder. The fluorescence exhibited by the sample is 
measured at right angles to the incident beam. The following is the basic set-up: 



Applications :  The applications of this technique are limited and it offers quantitative 
estimations of those compounds like benzene and fused benzene ring systems. Inorganic 
metals can also be analysed by the ability of  them to form complexes with the ligands. It 
finds uses in the analysis of foods for vitamin content, since vitamins like riboflavin, niacin, 
etc., exhibit fluorescence.

Only limited  compounds show the fluorescence hence this technique is relatively free of any interference and is very sensitive.

Disadvantages : The application is very limited as relatively a few substances exhibit
flourescence.



Ultraviolet Visible Spectroscopy UV/Vis

Ultraviolet - Visible Spectroscopy (UV/Vis): 

Principle :  It involves the absorption of electromagnetic radiation by the substances in the
ultraviolet and visible regions of the spectrum. This will result in changes in the electronic
structure of ions and molecules through the excitations of bonded and non-bonded electrons.

Instrumentation :  It consists of a dual light source viz., tungsten lamp for visible range and
deuterium lamp for ultraviolet region, grating monochromator, photo-detector, mirrors and
glass or quartz cells.

NOTE:  For measurements to be made under visible region both glass and quartz cells can be 
used. For the measurements under ultraviolet region, only quartz cell should be used, since, 
glass cells absorb ultraviolet rays.

There are two types of instrumental designs for this technique as single beam and double
beam spectrophotometers. However double beam spectrophotometers are widely used and
following is the outline of the instrument:




Applications :   It is the most widely used technique  for quantitative molecular analysis, for
this Beer-Lambert law is applied. Sometimes it is used in conjunction with other techniques
such as NMR, IR, etc., in the identification and structural analysis of organic compounds. For
qualitative analysis it provides valuable information through the absorption spectrum which
is unique for a given compound.

Disadvantages :  Samples should be in solution. Mixture of substances poses difficult to
analyse and requires prior separation. Interference from the sample’s matrix makes the
measurement difficult.


Atomic Fluorescence Fluorometry

Fluorometry : Atomic Fluorescence 
This technique is not widely used though its counterpart -  the molecular fluorescence is 
applied well to the analytical studies.



 The principle of atomic fluorescence is that when atoms are elevated to higher energy levels, they  sometimes return to the ground state through a 
pathway, which has several intermediate electronic states, before reaching to the actual 
ground state.
 Such series of fall through the electronic levels accompany by light emission - 
which is atomic fluorescence.
 The intensity of this emitted light is measured at right angles to 
the incident light and related to concentration.
Uses are similar to AAS and AES. 

Plasma Emission Spectroscopy

Plasma Emission Spectroscopy
Principle :  Mostly referred as Inductively Coupled Plasma (ICP) Emission Spectroscopy, is 
also an atomic emission technique, most closely related to the preceded flame photometry 
except that the atoms and ions present in the sample are excited in high temperature gas 
plasma.  Since the plasma provides very high  temperature and hence the energy, almost all 
the atoms present in the sample can be excited with this technique ending up with high 
efficiency (a hotter source increases both atomization  efficiency and excitation efficiency). 
Thus, the emissions from the atoms would  be more intense and even very small 
concentrations of metals/metal ions can be detected and accurately measured. 

Instrumentation:  This is basically an emission spectrometer comprising nebulizer, RF coil, 
ICP Source (Argon plasma), monochromator, detector and recorder. 



A plasma source or jet is a flame-like system of ionized, very hot flowing argon gas. At high 
temperatures (≈ 6000 K) a gas such as argon will contain a high proportion of ions and free 
electrons constituting plasma (This ionisation is initiated by “Tesla” coil). Additional energy 
may be supplied to the electrons in the plasma by the application of an external 
electromagnetic field through RF coil. By collisions between the electrons and other species 
in the plasma this additional energy is uniformly distributed. As the collisions increase, the 
energy transfer becomes more efficient, which leads to a substantial temperature 
enhancement to a range of 8000 - 10000 K. It is the temperature at which the samples are 
introduced and analysed. 

Applications : Similar to atomic emission spectroscopy but it covers very widespread for both 
qualitative and quantitative analysis of metals and some non-metals too, at trace levels. 
Because of the high temperature and homogeneity of the source, it offers better signal 
stability and hence the analytical precision. The technique when utilising an optical emission 
detector is termed as Inductively Coupled Plasma – Optical Emission Spectrometer  (ICP-
OES) and if it utilises a mass spectrometer (refer section 9.6) as detector then it is termed as 
Inductively Coupled Plasma – Mass Spectrometer (ICP-MS). 

Disadvantages:  Samples require dissolution before analysis. Instrumentation is complex and 
requires high operator’s skill and is very expensive. 

Atomic Emission Spectroscopy AES

Atomic Emission Spectroscopy (AES): 
Principle :  This is simply called as ‘Flame Photometry’, and measures the atoms excited by a
flame  (temperature range: 2000 – 31000
 K) and not by light source as in the atomic
absorption case. After excitation, atoms will readily lose the gained energy and revert back to
the ground state and the emission occurs. It is that emission that actually being measured.
The wavelengths of the emitted light will almost be similar as those that were absorbed in the
atomic absorption, since exactly the same energy transitions occur, except in the order of
reverse!    9
Instrumentation :  A simple flame photometer consists of burner, nebulizer, monochromator,
detector and recorder. The following is the simplified figure:




Applications :  It is used exclusively in  the quantitative determination of metals in solution,
especially alkali and alkaline earth in the given samples. The principle is like that described
for atomic absorption. Qualitative determination is also possible as  each element emits its
own characteristic line spectrum.

Disadvantages :  Intensity of emission is very sensitive to changes in flame temperature.
Usually, spectral interference and self-absorption are also  encountered which affects the
precision of the measurement. Further, a linear plot of absorbance against concentration is
not always obtained.

Atomic Absorption Spectroscopy AAS

Atomic Level 
Atomic Absorption Spectroscopy  AAS
Principle: The sample is vaporized by aspiration of solution into a flame or evaporation from 
electrically heated surface (temperature range: 1800 – 31000
 K). At this condition where the 
individual atoms co-exist, a beam of light is passed through them. The atoms will absorb in 
the visible and ultraviolet region resulting in changes in electronic structure (excited state). 
So, the resultant light beam coming out of the sample will be missing the light in the 
corresponding wave length, which is a measure of the characteristics of the sample. 

Instrumentation :  Sources emitting radiation characteristic of element of interest (hollow - 
cathode lamp), flame or electrically heated furnace, monochromator, detector 
(photomultiplier) and  recorder. The following is the simplified outline of the 
instrumentation: 




Applications : This is the most widely used technique for the quantitative determination of 
metals at trace levels (0.1 to 100ppm), which present in various materials. It utilizes Beer  - 
Lambert Law  for the analysis and a standard curve is obtained by plotting absorbance  vs 
concentration of the samples taken. The usual procedure is to prepare a series of standard 
solutions over a concentration  range suitable for the sample to be analysed. Then, the 
standards and the samples are separately aspirated into the flame, and the absorbances are 
read from the instrument. The plot will give the useful linear range and the concentrations of 
the samples can be found out from the plot. 

Disadvantages : Sample must be in solution or at least volatile. Individual source lamp and 
filters needed for each element, since, each metal has its own characteristic absorption.

Processes in Spectroscopy


ANALYSIS THROUGH SPECTROSCOPY 
 Processes in Spectroscopy 
The interaction of the light (electro-magnetic radiation) with a substance and the subsequent 
energy transfer ends with three main processes namely: 

Absorption:
The process by which the energy of the light (in the form of photons) is transferred to the 
atom or molecule raising them from the ground state to an excited state 

Fluorescence: 
The absorbed energy is rapidly lost to the surroundings by collisions within the system and 
relax back to the ground state. Sometimes the energy is not lost in this way but is re-emitted a 
few milli seconds later, which is referred as fluorescence 

Emission:          
If the substances (atoms or molecules) are heated to high temperatures (in a flame or in an 
electric discharge) the electrons are exited to higher energy levels. Later, they relax to the 
ground state with the emission of radiation, the magnitude of which is more or less equivalent 
to absorbed energy   

Most of the analytical techniques are based on the light interactions with the substances and 
utilise any of the above three associated processes. Substances interact with light differently 
at various wavelengths and hence different types of analysis & instruments. The entire 
spectrum of light can be represented as below. Since, light has both electrical and magnetic 
components, this representation is referred as an ‘Electro-Magnetic Spectrum’: 







The following is short comparison between Ultra Violet (UV), Visible (Vis) and Infra Red  
(IR) ranges for the energy, frequency and wavelength:

Energy:   UV > Vis > IR 
Frequency:   UV > Vis > IR 
Wavelength:   UV < Vis < IR         

The symbol for the wavelength is  “lambda” (λ) and the unit is either nanometer (nm) or 
micrometer (or micron, μm). The symbol for frequency is “nu” (μ) and the unit is either hertz 
or sec-1
A parameter closely related to frequency is the wave number, which has the symbol 
“nu bar” ( υ ) and the unit is cm-1

There are two levels by which the substances can interact with the light as, atomic level and 
molecular level and hence the corresponding techniques: 

CLASSIFICATION OF THE ANALYTICAL TECHNIQUES

CLASSIFICATION OF THE ANALYTICAL TECHNIQUES 

In a broad sense the techniques for the chemical analysis can be classified as follows. Though 
this classification doesn’t include few other techniques like radiochemical analyses, 
bioanalytical methods and some of the physical methods, it is more than sufficient to start 
with, since it covers almost all our Departmental analytical equipment under common pool:  


ANALYSIS THROUGH SPECTROSCOPY 

ANALYSIS THROUGH CHROMATOGRAPHY 

ANALYSIS THROUGH THERMAL ENERGY 

ANALYSIS THROUGH X-RAY TECHNIQUES 

ANALYSIS THROUGH MICROSCOPY  

ANALYSIS THROUGH ELECTROCHEMICAL TECHNIQUES 

ANALYSIS THROUGH MISCELLANEOUS TECHNIQUES 

This classification is based on the interactions of molecules with various forms of energy like 
electro-magnetic radiation, heat (thermal energy) and with  matters like electrons. Each 
technique has specific principle, mode of operation, advantages and disadvantages. 

Monday, January 2, 2012

anharmonic oscillator spectrum


anharmonic oscillator spectrum [SPECT] A molecular spectrum which is significantly
affected by anharmonicity of the forces between atoms in the molecule

angle-resolved photoelectron spectroscopy


angle-resolved photoelectron spectroscopy [SPECT] A type of photoelectron spec
troscopy which measures the kinetic energies of photoelectrons emitted from a solid
surface and the angles at which they are emitted relative to the surface. Abbreviated
ARPES

Thursday, December 29, 2011

analysis line


analysis line [SPECT] The spectral line used in determining the concentration of an
element in spectrographic analysis

Monday, July 4, 2011

air line

air line [SPECT] Lines in a spectrum due to the excitation of air molecules by spark
discharges, and not ordinarily present in arc discharges.

Saturday, June 18, 2011

accelerator mass spectrometer

accelerator mass spectrometer [SPECT] A combination of a mass spectrometer and
an accelerator that can be used to measure the natural abundances of very rare
radioactive isotopes.

Friday, June 17, 2011

absorption spectrum

absorption spectrum [SPECT] A plot of how much radiation a sample absorbs over a
range of wavelengths; the spectrum can be a plot of either absorbance or transmit-
tance versus wavelength, frequency, or wavenumber. { əb so ˙ rp shən spek trəm}

absorption spectroscopy

absorption spectroscopy [SPECT] An instrumental technique for determining the con-
centration and structure of a substance bymeasuring the intensity of electromagnetic
radiation it absorbs at various wavelengths. { əb so ˙ rp shən spek tra ¨s kə pe