Fourier Transform Infrared Spectroscopy FT-IR

Fourier Transform Infrared Spectroscopy (FT-IR): 
Principles :   It involves the absorption of electromagnetic radiation in the infrared region of 
the spectrum which results in changes in the vibrational energy of molecule. Since, usually 
all molecules will be having vibrations in the form of stretching, bending, etc., the absorbed 
energy will be utilised in changing the energy levels associated with them. It is a valuable 
and formidable tool in identifying organic compounds which have polar chemical bonds 
(such as OH, NH, CH, etc.) with good charge separation (strong dipoles).  

Instrumentation :  It was originally designed as a double beam spectrophotometer comprising 
IR source (red hot ceramic material), grating monochromator, thermocouple detector, cells 
made of either sodium chloride or potassium bromide materials, etc. In this process the light 
is dispersed by the monochromator. But, this type of basic design for IR measurements has 
been outdated. Instead a newer technique termed Fourier Transform-Infrared (FT-IR) has 
been in practice. This technique utilises a single beam of un-dispersed light and has the 
instrument components similar to the previous one. 

In FT-IR, the un-dispersed light beam is passed through the sample and the absorbances at all 
wavelengths are received at the detector simultaneously. A computerized mathematical 
manipulation (known as “Fourier Transform”) is performed on this data, to obtain absorption 
data for each and every wavelength. To perform this type of calculations interference of light 
pattern is required for which the FT-IR instrumentation contains two mirrors, one fixed and 
one moveable with a beam splitter in between them. Before scanning the sample a reference 
or a blank scanning is required. The following is the simplified design of the instrument: 




Applications :  It finds extensive use in the identification and structural analysis of organic 
compounds, natural products, polymers, etc. The presence of particular functional group in a 
given organic compound can be identified.  Since every functional group has unique 
vibrational energy, the IR spectra can be seen as their fingerprints. 

Disadvantages :  Samples containing mixture of substances can not be analysed. Since the 
sample holders and beam splitter, are made  of moisture sensitive materials like sodium 
chloride or potassium bromide (KBr), special cells are required for aqueous samples (e.g. 
KRS-5, ZnSe, etc.). Water is a bad solvent for IR spectral works.

Fluorometry Molecular Fluorescence


Fluorometry:  Molecular Fluorescence 
Principle :  This technique utilises the phenomenon  of molecular fluorescence, the theory 
behind this is exactly the same that has been discussed under atomic fluorescence but through 
the excitation of bonded electrons. Here, most often the irradiating light is in the range of 
ultraviolet and visible.  

Instrumentation :  The instrumental set-up comprises of a UV/Visible source, two 
monochromators, detector and recorder. The fluorescence exhibited by the sample is 
measured at right angles to the incident beam. The following is the basic set-up: 



Applications :  The applications of this technique are limited and it offers quantitative 
estimations of those compounds like benzene and fused benzene ring systems. Inorganic 
metals can also be analysed by the ability of  them to form complexes with the ligands. It 
finds uses in the analysis of foods for vitamin content, since vitamins like riboflavin, niacin, 
etc., exhibit fluorescence.

Only limited  compounds show the fluorescence hence this technique is relatively free of any interference and is very sensitive.

Disadvantages : The application is very limited as relatively a few substances exhibit
flourescence.



Ultraviolet Visible Spectroscopy UV/Vis

Ultraviolet - Visible Spectroscopy (UV/Vis): 

Principle :  It involves the absorption of electromagnetic radiation by the substances in the
ultraviolet and visible regions of the spectrum. This will result in changes in the electronic
structure of ions and molecules through the excitations of bonded and non-bonded electrons.

Instrumentation :  It consists of a dual light source viz., tungsten lamp for visible range and
deuterium lamp for ultraviolet region, grating monochromator, photo-detector, mirrors and
glass or quartz cells.

NOTE:  For measurements to be made under visible region both glass and quartz cells can be 
used. For the measurements under ultraviolet region, only quartz cell should be used, since, 
glass cells absorb ultraviolet rays.

There are two types of instrumental designs for this technique as single beam and double
beam spectrophotometers. However double beam spectrophotometers are widely used and
following is the outline of the instrument:




Applications :   It is the most widely used technique  for quantitative molecular analysis, for
this Beer-Lambert law is applied. Sometimes it is used in conjunction with other techniques
such as NMR, IR, etc., in the identification and structural analysis of organic compounds. For
qualitative analysis it provides valuable information through the absorption spectrum which
is unique for a given compound.

Disadvantages :  Samples should be in solution. Mixture of substances poses difficult to
analyse and requires prior separation. Interference from the sample’s matrix makes the
measurement difficult.


Atomic Fluorescence Fluorometry

Fluorometry : Atomic Fluorescence 
This technique is not widely used though its counterpart -  the molecular fluorescence is 
applied well to the analytical studies.



 The principle of atomic fluorescence is that when atoms are elevated to higher energy levels, they  sometimes return to the ground state through a 
pathway, which has several intermediate electronic states, before reaching to the actual 
ground state.
 Such series of fall through the electronic levels accompany by light emission - 
which is atomic fluorescence.
 The intensity of this emitted light is measured at right angles to 
the incident light and related to concentration.
Uses are similar to AAS and AES. 

Plasma Emission Spectroscopy

Plasma Emission Spectroscopy : 
Principle :  Mostly referred as Inductively Coupled Plasma (ICP) Emission Spectroscopy, is 
also an atomic emission technique, most closely related to the preceded flame photometry 
except that the atoms and ions present in the sample are excited in high temperature gas 
plasma.  Since the plasma provides very high  temperature and hence the energy, almost all 
the atoms present in the sample can be excited with this technique ending up with high 
efficiency (a hotter source increases both atomization  efficiency and excitation efficiency). 
Thus, the emissions from the atoms would  be more intense and even very small 
concentrations of metals/metal ions can be detected and accurately measured. 

Instrumentation:  This is basically an emission spectrometer comprising nebulizer, RF coil, 
ICP Source (Argon plasma), monochromator, detector and recorder. 



A plasma source or jet is a flame-like system of ionized, very hot flowing argon gas. At high 
temperatures (≈ 6000 K) a gas such as argon will contain a high proportion of ions and free 
electrons constituting plasma (This ionisation is initiated by “Tesla” coil). Additional energy 
may be supplied to the electrons in the plasma by the application of an external 
electromagnetic field through RF coil. By collisions between the electrons and other species 
in the plasma this additional energy is uniformly distributed. As the collisions increase, the 
energy transfer becomes more efficient, which leads to a substantial temperature 
enhancement to a range of 8000 - 10000 K. It is the temperature at which the samples are 
introduced and analysed. 

Applications : Similar to atomic emission spectroscopy but it covers very widespread for both 
qualitative and quantitative analysis of metals and some non-metals too, at trace levels. 
Because of the high temperature and homogeneity of the source, it offers better signal 
stability and hence the analytical precision. The technique when utilising an optical emission 
detector is termed as Inductively Coupled Plasma – Optical Emission Spectrometer  (ICP-
OES) and if it utilises a mass spectrometer (refer section 9.6) as detector then it is termed as 
Inductively Coupled Plasma – Mass Spectrometer (ICP-MS). 

Disadvantages:  Samples require dissolution before analysis. Instrumentation is complex and 
requires high operator’s skill and is very expensive. 

Atomic Emission Spectroscopy AES

Atomic Emission Spectroscopy (AES): 
Principle :  This is simply called as ‘Flame Photometry’, and measures the atoms excited by a
flame  (temperature range: 2000 – 31000
 K) and not by light source as in the atomic
absorption case. After excitation, atoms will readily lose the gained energy and revert back to
the ground state and the emission occurs. It is that emission that actually being measured.
The wavelengths of the emitted light will almost be similar as those that were absorbed in the
atomic absorption, since exactly the same energy transitions occur, except in the order of
reverse!    9
Instrumentation :  A simple flame photometer consists of burner, nebulizer, monochromator,
detector and recorder. The following is the simplified figure:




Applications :  It is used exclusively in  the quantitative determination of metals in solution,
especially alkali and alkaline earth in the given samples. The principle is like that described
for atomic absorption. Qualitative determination is also possible as  each element emits its
own characteristic line spectrum.

Disadvantages :  Intensity of emission is very sensitive to changes in flame temperature.
Usually, spectral interference and self-absorption are also  encountered which affects the
precision of the measurement. Further, a linear plot of absorbance against concentration is
not always obtained.